Photonic Lantern Digital Holography Automation

Automated measurement and analysis of the wavelength-dependent transfer matrix of a photonic lantern by off-axis digital holography — for Prof. Stephen Eikenberry's group at UCF CREOL.

A photonic lantern maps several single-mode fibers to the guided modes of a multimode fiber; that mapping is a complex-valued transfer matrix that evolves with wavelength. This software automates measurement of the encoding matrix: it sweeps the input port and wavelength, captures an off-axis hologram of the multimode output at each point, and recovers the complex field (amplitude and phase) for decomposition into LP modes — one row of the transfer matrix per measurement.

The reconstruction follows the method in:

C. Dobias, M. A. Römer, S. Bhargava, et al., "Wavelength-dependent evolution of full-field transfer matrices in photonic lanterns," Optics Express 34(9), 17217 (2026). https://doi.org/10.1364/OE.595908


What it does

  • Acquire — for each lantern port (fiber switch) × wavelength (tunable laser), auto-optimize polarization for maximum fringe contrast, then capture the off-axis hologram. Frames are validated (saturation/clipping is rejected).
  • Reconstruct — FFT the hologram → isolate the off-axis sideband → demodulate to baseband → Butterworth low-pass → inverse FFT to the complex field; optimize mode-field diameter, quadratic-phase correction, and position for best fidelity; decompose into LP modes.
  • Run unattended — the goal is a full 1525–1575 nm × all-ports sweep with only a couple of minutes of human time: line it up, hit start, walk away.

Hardware

Component Model Interface
Tunable laser HP/Agilent 8168E (1475–1575 nm) GPIB · Keysight VISA
Camera Xenics Bobcat 320 GigE (InGaAs, 320×256) GigE Vision
Fiber switch DiCon GP700 RS-232
Polarization Thorlabs MPC320 (3 paddles) USB · Kinesis

Quick start

The shipping app is a single native Windows executable — no Python, Node, server, or browser needed on the lab machine:

uv run python tools/build_app.py      # -> dist/Digital Holography.exe

Copy dist/Digital Holography.exe to the lab PC and double-click. Connect All → Start Experiment. Config, logs, and data are written next to the .exe.

Run from source instead (needs uv; no Node):

uv sync
./start.bat            # = uv run python main.py

One-time per machine: install Keysight IO Libraries + Xenics Xeneth SDK and run the camera firewall rule — see LAB_SETUP.md.

Repository layout

holo/ is the library — every line of physics lives there. The desktop app and the holo command are two front ends over it, and neither contains analysis code of its own, so they can never drift apart.

holo/                          THE LIBRARY
  pipeline.py                  combines both engines — change processing here
  data_processing.py           single-frame reconstruction + LP decomposition
  multiport_reconstruction.py  cross-port (paper-fidelity) reconstruction + TM
  fringe_detection.py          fringe metric + polarization auto-optimizer
  discovery.py                 folder -> records; frame loading and staging
  config.py                    config loading
  runtime.py                   frozen-build paths + Xeneth DLL bootstrap
  cli.py                       the `holo` command
  lib/                         LP modes, FFT math, vendored Xeneth SDK
  hardware/                    drivers (laser, camera, switch, motors)

gui/                        PySide6 (Qt6) app — window, tabs, acquisition loop
main.py                     GUI launcher (thin entry point)
tools/                      build script + hardware probes
tests/                      golden-file + unit tests (`uv run pytest`)
experiment_config.yaml      instrument addresses + sweep parameters
docs/                       setup guides (GigE camera, Xeneth SDK)

Command line

holo acquire -o ./run            # sweep 1530-1560 nm, 3 exposures per point
holo process ./data              # reconstruct a folder of holograms
holo tm ./data -o tm.npz         # reconstruct + export the transfer matrix
holo doctor                      # check the install, config and instruments
holo probe switch                # talk to one instrument
holo gui                         # launch the desktop app

uv run holo <command>, or python -m holo without installing. Every command takes --help. process.py still works as a deprecated shim for holo process.

Acquiring a sweep by hand

holo acquire drives only the laser and the camera — you connect the lantern leg and set the paddles yourself, which is the usual way to take a single-leg test. Everything is a flag; no config file to edit.

holo acquire --start 1530 --stop 1560 --step 1 -n 3 --leg 1 -o ./run
holo acquire --start 1530 --stop 1560 -n 3 --dry-run   # print the plan, no hardware

-n 3 takes three exposures back to back at each wavelength and writes them to run/rep01, run/rep02, run/rep03. They are separate folders on purpose: frame discovery keeps one file per (leg, wavelength), so repeats sharing a folder would leave all but one silently unprocessed. Each repNN is a complete sweep — reconstruct one with holo process ./run/rep01.

Every frame gets a .png preview and a .yaml sidecar recording the commanded and measured wavelength (the HP 8168E queues errors instead of raising, so a rejected tune otherwise looks like a good frame at the wrong λ). Saturated frames are flagged and the command exits non-zero — clipping makes intensity nonlinear in the field, so both amplitude and phase are wrong.

Using it as a library:

from holo import HolographyDataProcessor, pipeline, discovery

records, cache = discovery.stage("./data", default_wl=1550)
rows = pipeline.process_records(HolographyDataProcessor(), records, config=cfg)

See CLAUDE.md for how the two reconstruction engines relate and where the physics lives.

Reconstruction

Two engines, picked automatically by how much data the sweep has:

single-frame multiport
module data_processing.py multiport_reconstruction.py
needs one hologram ≥ 2 legs across the sweep
carrier per-frame sub-pixel centroid, demodulated to baseband centroid averaged over all ports per λ, then linear-fit vs λ, then Fourier Fine-Binning
basis per-frame optimum optima consolidated across the sweep (paper Sec. 2)

Both compute the paper's fidelity η = |⟨E_rec, E_S⟩ / (‖E_rec‖‖E_S‖)|² (Eq. 5). A multi-leg dataset runs both and keeps, per frame, whichever scored higher — so the result can never be worse than single-frame.

Status & known limits

  • Camera streaming, laser, motors, acquisition, the fringe auto-optimizer, and saturation rejection all work.

  • Single-frame reconstruction reaches 0.9717 on the committed test frame (15-mode basis; the 8-mode basis gives 0.9497 on the same frame). Pixel scale, mode-field diameter, field position, and the defocus quadratic-phase are optimized numerically (matching the paper — they are not measured on the bench).

  • That number is not directly comparable to the paper's 98%. The paper measures a 19-port lantern with a 23-mode basis; this config uses a 15-mode basis on what the multiport path treats as a 7-port (8-mode) lantern. An oversized basis inflates fidelity by absorbing noise — see CLAUDE.md, "Two open physics questions". Resolve the lantern's real mode count before quoting a fidelity against the paper.

  • Multiport is a verified reproduction of Caleb's analysis. Carrier centroids identical to 0.000 px, an identical interpolated twin image (|overlap| = 1.000000), and mode decompositions matching to 0.997–0.9998 cosine similarity. On fidelity, quote the scope with the number:

    consolidated run frames mean σ worst < 95%
    3 λ (1525/1550/1575) × 19 legs 57 96.76% 0.95% 93.7% 1
    full 51 λ × 19 legs 969 96.68% 1.58% 80.0% 37

    The mean is stable across both (96.7%), so the reproduction holds. The spread is not: over the full sweep σ is 1.58% and 37 frames fall below 95% (10 below 90%). Earlier revisions of this file quoted the 57-frame row as if it covered the whole archive and concluded the consolidated basis beat Caleb's worst frame (81.1%, 24 under 95%). On the full sweep it does not — 80.0% is his worst, near enough. Whether his figures span 57 or 969 frames is unrecorded, so treat that comparison as open rather than won.

  • The paper's printed 98% ± 0.8% is not reproducible from that dataset — not here and not by Caleb's own saved run, which also lands at 96.9%. The ~1.1 point difference sits between his 2023 analysis and the published figure. See CLAUDE.md, which also records why tuning the Butterworth passband to hit 98% is a false fix. Pinned by tests/test_multiport.py (auto-skips without the dataset, which is gitignored and multi-GB).

  • Multiport has not yet been validated on a leg × wavelength sweep from this rig — only on the paper's data. On the rig data available today it underperforms and the pipeline falls back to single-frame per frame.

  • The DiCon GP700 switch is being replaced — its driver may need updating for the new unit.